Product introduction
AUTO W46/1 Infrared Method Water Vapor Transmission Rate Tester is based on the testing principle of the infrared method, designed with reference to GB/T 26253, ASTM F1249 and other standards, and tests the water vapor transmission rate of samples under set temperature and humidity conditions (quantity), suitable for water vapor transmission rate performance testing of films, sheets, paper, packages and various materials in the fields of food, medicine, medical equipment, daily chemicals, photovoltaic electronics, etc. It is an ideal configuration instrument for off-line or on-line testing of the barrier properties of packaging materials for production units in the packaging industry.
Test principle
AUTO W46/1 water vapor transmission rate tester adopts the principle of infrared method. Fix the pre-treated sample in the middle of the test chamber, divide the test chamber into a high-humidity side and a low-humidity side, the humidified nitrogen flows on one side of the film, and the dry nitrogen (carrier gas) flows at a fixed flow rate on the other side, in the presence of a humidity gradient, water vapor will permeate from the high-humidity side to the low-humidity side, and the water vapor that passes through the sample is carried to the infrared sensor by the flowing dry nitrogen, and the water vapor transmission rate (and other parameters) of the sample is obtained by the electrical signal output by the sensor.
Standards
ASTM F1249, GB/T 26253, YBB 00092003, BS EN ISO 15106-2, JIS K7129
Technical Parameters
Item |
Technical parameters |
Test range |
0.002~200 g/(m²·24h)(film and sheet) |
Test precision |
0.0001 g/(m²·24h)(film and sheet) |
Temperature range |
15~45℃(15~60℃ optional) |
Temperature accuracy |
±0.1℃ |
Humidity range |
(5~90)%RH,100%RH |
Humidity accuracy |
±2%RH |
Test area |
50.24 cm² |
Sample size |
Φ110 mm |
Sample thickness |
≤3 mm |
Number of test sample |
6 Pieces |
Carrier gas |
99.999% N2(user provide) |
Carrier gas pressure |
≥0.1 MPa |
Carrier gas flow |
5~100 mL/min |
Pneumatic pressure |
≥0.3 MPa |
Gas supply port |
1/8 inch metal pipe |
Instrument size |
460 mm×592 mm×628 mm |
Power |
1150 W |
Power supply |
AC 220 V,50 Hz (110V as requested) |
Features
Unique cavity design, innovative and convenient to upgrade
The newly upgraded vertical drawer cavity design can be opened and closed with one button, which is convenient for upgrading.
The fully automatic pneumatic clamp clamps the sample, which is convenient and labor-saving, and has excellent sealing performance.
High-precision patented infrared moisture sensor
The high-precision infrared moisture sensor developed and prepared by the top technical team at home and abroad has ultra-high stability and ultra-low failure rate, long service life, high sensitivity, and a resolution of up to 0.0001 g/(m²·24h).
Precise control of temperature and humidity
Temperature control: bidirectional automatic temperature control of the semiconductor refrigeration chip, the temperature control accuracy reaches 0.1 ℃.
Humidity control: dual air flow (dry gas and wet gas) humidity control method, stable humidity, high precision, accurate humidity to ±2%RH.
Meet the test requirements of high throughput and wide range
The instrument is equipped with 6 chambers with independent data, which can meet the needs of high-throughput testing and has high testing efficiency.
The measuring range is 0.002~200 g/(m²·24h), and the measuring range is wide, meeting the testing requirements of high, medium and low barrier materials.
Application field
|
Films |
Water vapor transmittance test of various plastic films (PP / PET / PE / PVC / BOPP / CPP, etc.), plastic composite films, paper plastic composite films, metal composite films, co-extrusion films, aluminum films, degradable packaging films (PLA / PBAT / PBS, etc.) |
|
Sheet |
Water vapor transmission rate test of solid pharmaceutical hard sheet (PP/PVC/PTP, etc.), metal composite sheet, rubber sheet and other sheet materials |
|
Paper, cardboard and their composites |
Water vapor transmission rate test of paper and cardboard such as coated paper, silicone paper, aluminum paper for cigarette packs, paper-aluminum-plastic composite sheet, etc. |
|
Medicinal patch |
Water vapor transmission performance test of medical plasters |
|
Package |
Customized fixtures can be extended to packaging parts, such as the water vapor transmission rate of medical polyethylene bottles, sealed bags, medical ointment tubes, infusion hoses, plastic trays, etc. |