Standards
YY0469-2004、YY/T 0969-2013、ASTMF2100、ASTMF2101、EN14683
Specifications:
Key Specification
|
Specification Range |
Resolution |
Accuracy |
A Route Sampling Flow |
28.3L/min |
0.1L/min |
Within ± 2.5% |
B Route Sampling Flow |
28.3L/min |
0.1L/min |
Within ±2.5% |
Spray Flow |
(8~10)L/min |
0.1L/min |
Within ±2.5% |
Peristaltic Pump Flow |
(0.006~3.0)mL/min |
0.001ml/min |
Within ±2.5% |
Front Pressure of A Route Sampling Flowmeter |
(-20~0)kPa |
0.01kPa |
Within ±2.5% |
Front Pressure of B Route Sampling Flowmeter |
(-20~0)kPa |
0.01kPa |
Within ±2.5% |
Front Pressure of Spray Flowmeter |
(0~300)kPa |
0.1kPa |
Within ±2.5% |
Ambient Temperature |
(-40~99)℃ |
0.1℃ |
Within ±2.5% |
Negative Pressure of the Aerosol Chamber |
(-90~-120)Pa |
0.1Pa |
Within ±2.0% |
Negative Pressure of the Cabinet
|
-50~-200pa |
||
Data Storage Ability |
>100000 sets |
||
Vortex Mixer Test Tube Specification and Quantity
|
Φ16×150mm test tube,8pcs
|
||
High Efficiency Particulate Air Filter Properties |
Filter Efficiency≥99.99% for particles> 0.3μm |
||
Median diameter of aerosol generator mass |
Average diameter: (3.0±0.3)μm; Geometric Standard Deviation≤1.5 |
||
Double Routes 6-lever Andersen Sampling apparatus |
Level1>7μm; Level 2: (4.7~7)μm; Level 3: (3.3~4.7)μm; Level 4: (2.1~3.3)μm; Level 5: (1.1~2.1)μm; Level 6: (0.6~1.1)μm |
||
Size of Aerosol Chamber
|
600×85×3mm (Length×Diameter×Thickness) |
||
Total number of positive quality control sampler particles
|
(2200±500)cfu |
||
Negative Chamber Flow Speed |
≥5m3/min |
||
Size of Negative Pressure Chamber Door(W×D) |
1000×730mm |
||
Size of Main Machine(W×D×H) |
1180×650×1300mm |
||
Size of Supporting Frame(W×D×H) |
1180×650×600mm, Height within 10cm is adjustable |
||
Noise of the Tester |
<65dB(A) |
||
Weight
|
150kg |
||
Power Supply |
AC220V±10%, 50Hz |
||
Power Consumption
|
<1500W |
Features
1. Negative pressure experiment system to ensure the safety of operators;
2. Negative pressure cabinet with built-in peristaltic pump and A and B two-way six-stage Andersen;
3. The flow rate of the peristaltic pump can be set;
4. Dedicated microbial aerosol generator can set the volume of bacterial liquid spray flow, and the atomization effect is good;
5. Embedded high-speed industrial microcomputer control;
6.10.4-inch industrial-grade high-brightness color touch display;
7, USB interface, support U disk data transfer;
8. Built-in high-brightness lighting in the cabinet;
9. Built-in leakage protection switch to protect operator safety;
10. Insulation and flame retardant between inner and outer layers;
11. The front switch glass door is convenient for experimenters to observe and operate;
12. Detachable stand, the height of stand is adjustable;
13. Support and move dual-purpose casters.