- Salt spray test chamber GB-OSH-60View More
1. FunFunction
Salt spray test chamber is used for the quality inspection and corrosion resistance test of a variety of materials, including metal plating, metal, electronic components, chemical coatings, paint, automobiles, motorcycles, metal ware, screws, springs, magnetic materials, organic and inorganic coating, anodizing, anti-rust oil industry etc. after the surface treatment.
1. ecGTechnical parameters
1. MODEL
GB-OSH-60
2. Temperature inside chamber
35℃±1℃/ 50℃±1℃
3. Temperature of pressure barrel
47℃±1℃/ 63℃±1℃
4. Inner size
W600 x H400 x D450mm
5. External size
W1080 x H1080 x D650mm
6. Temperature Uniformity
≤±2℃
7. Uniformity Fluctuation
≤±0.5℃
8. Salt fog settlement
1~2ml/h.80cm2
9. Compressed air pressure
1.00±0.01kgf/cm²
10. Spray mode
Continuous or intermittent spray optional
11. Power supply
AC 220V 50Hz 1.5KW
a) Temp. range: RT+10℃ ~ 55℃ (temperature up to 50C, thickening PVC or PP material is recommended )
b) Humid. range: ≥93%R.H
c) Controller time shown: 9.9s~9990hrs
d) Timer: 0~9999hrs.
e) Testing method: Neutral salt spray test (NSS test), acetic acid salt spray test (AASS test), salt spray test (SS test), copper accelerated acetic acid salt spray test method for salt spray test (CASS test) and so on; (For CASS test, we recommend the PP material chamber with high temperature resistance)
f) Specimen holder: It can meet the tilt test at 15 ℃ ~ 30 ℃ (two layers);
g) Air compressor: V-0.013/12.5(Customer provided); discharge pressure:1.25M; flow amount: 0.13m3/min
1. MMMaterial of machine body
a) Salt spray chamber adopts the quality PVC reinforced hard plastic plate, the surface is smooth and anti-aging, anti-corrosion;
b) Without any FRP material produced on the surface because of discoloration fibrosis. Plastic structure of the chamber is better to meet strong acid salt spray test under long-term without any damage.
c) Cover is imported transparent impact-resistant plate which is convenient to test specimen when test, that can cover and the box with a water seal to prevent leakage of salt spray.
2. AccAccording to CSN, JIS, ASTM standards, can be set to constant temperature control:
(a) Brine spray test: NSS (neutral);
· Temp. inside chamber: 35 ℃ ± 1 ℃
· Pressure air bucket: 47 ℃ ± 1 ℃
(b) Corrosion resistance test: CASS (copper acceleration), AASS (acid)
· Temp. inside chamber: 50 ° C ± 1 ° C
· Pressure air bucket: 63 ℃ ± 1 ℃
3. Watwatering system.
It adopts the auto watering system, supply water when lack of water.
4. HeaHeating system
Test temperature inside the cabinets is heated humidification water jacket, the heater adopts advanced titanium protective tube, built-in nickel-chromium alloy infrared heater core. Heating up fast, uniform temperature distribution; and saves about one half of the power compared to FRP tank jacketed heating.
5. ExeExecutive standard
GB/T 2423.17 Basic test procedures for electric and electronic products Ka"- salt spray test methods;
GB/T 10125 Corrosion tests in artificial atmospheres – salt spray tests;
ASTM B368 Test Method for Copper-Accelerated Acetic Acid-Salt Spray (Fog) Testing (CASS Test);
JIS Z2371-2000 Methods of salt spray testing etc.
ISO 9227 Corrosion tests in artificial atmospheres (Note: 2 collection device is required, bigger models than ZH-SH-90 (270L) is required; In some versions, it specifies over 400liters is necessary)
ASTM B117-18 Standard Practice for Operating Salt Spray (Fog) Apparatus (Note: 2 collection device is required, bigger models than ZH-SH-90 is required )